Connection Guide for Wafer Prober and Your Own Test Fixture
To Connect HCSMU Adapter
The following connection example uses one HCSMU module and the 16493S-010 adapter
for making the Kelvin connection. In this example, the GNDU path is prepared as a
separate path. This GNDU connection is also effective for the connection using the
16493S-011 non-Kelvin adapter.
To simplify the GNDU connection, change the connection as shown in
connection can save one manipulator.
Figure 5-3
Typical Connection to Use One HCSMU, Kelvin connection
B 1505A
Figure 5-4
To Simplify GNDU Connection, Kelvin connection
Figure 5-5
To Simplify GNDU Connection, non-Kelvin connection
5-18
16493J
16493S
16493L
16493S-010
GNDU adapter (N1261A-002)
Low Sense
Low Force
16493S-010
*2
Force
Sense
GNDU adapter (N1261A-002)
Low Force
16493S-011
*2
Force
Sense
GNDU adapter (N1261A-002)
*1: 1252-1419
*2: N1254A-503 or other cable, 5 ea.
*3: N1254A-516, 1 ea. and 16493U-001, 2 ea.
*1
Interlock circuit
*2
High
Low
*3
*1
*1: N1254A-503 or other cable, 2 ea.
*2: 1250-2405, 2 ea. and 8120-1838 or 1839, 2 ea.
*1
*1: N1254A-503 or other cable, 1 ea.
*2: 1250-2405, 2 ea. and 8120-1838 or 1839, 2 ea.
Keysight B1505A Configuration and Connection Guide
Figure
5-4. This
Shielding box
Force
DUT interface
Sense
Low
High
Sense
*2
Force
*2
To DUT interface
To DUT interface