N1265A Connection Guide
Input Connection
Table 3-4
To Connect between B1505A and N1265A for using UHVU (N1268A)
B1505A terminals
16493J-001 Interlock cable, 1.5 m, 1 ea.
Interlock
MCSMU
Force
16494A-001 Triaxial cable, 1.5 m, 1 ea.
Sense
16494A-001 Triaxial cable, 1.5 m, 1 ea.
2
Force
16494A-001 Triaxial cable, 1.5 m, 1 ea.
MCSMU
Sense
16494A-001 Triaxial cable, 1.5 m, 1 ea.
1. To configure UHVU, the N1268A ultra high voltage expander is required.
2. HCSMU is substitutable. Then use 16493S HCSMU cable, N1254A-103 adapter for Force, and
N1254A-517 adapter for Sense.
N1268A terminals
16493J-001 Interlock cable, 1.5 m, 1 ea.
Interlock Output
Output
High
UHV(plug)-UHV(plug) cable, 1.5 m, 1 ea. (part of 16493V-001)
Low
SHV(plug)-SHV(plug) cable, 1.5 m, 1 ea. (part of 16493V-001)
16493G-001 Digital I/O connection cable, 1.5 m, 1 ea.
1
Digital I/O
LINE
Power cable, 1 ea., for connecting to power line
(earth)
Wire, 1 ea., for earthing
1. If the N1266A expander is used, connect the 16493G-001 cable from the N1266A Digital I/O Output.
WARNING
Connect a wire from an electrical ground (safety ground) to the earth terminal.
3-8
Required cables
Required cables
Keysight B1505A Configuration and Connection Guide
1
N1268A terminals
Interlock Input
Force
I Control
Sense
Force
V Control
Sense
N1265A terminals
Interlock
High
UHV
Low
Digital I/O Output