N1259A Connection Guide
Output Connection

N1259A-013 Curve Tracer Test Adapter Socket Module

This module provides a socket available for connecting a test adapter designed for
connecting to Tektronix 370 series curve tracers. Socket module internal connection is
shown in
To use this module, see the following simple instruction.
CAUTION
Do not apply voltage/current over the maximum limit of the socket module.
Figure 2-13
Curve Tracer Test Adapter Socket Module
2-22
Figure
2-13.
Required parts:
N1254A-508 or N1254A-509 connection wire, 1 ea. for one connection
Instruction:
1. Attach the socket module to the test fixture.
2. Connect your test adapter to the socket.
3. Connect wires between the socket module terminals and the fixture output
terminals.
For making the Kelvin connection, Force and Sense must be connected to Force
and Sense of the socket module respectively.
If MCSMU, HCSMU, or DHCSMU is used, High Force and Sense must be
connected to Force and Sense used for the high terminal of a device under test
(DUT), respectively. Low Force and GNDU Force must be connected to Force used
for the low terminal of the DUT, and Low Sense and GNDU Sense must be
connected to Sense used for the low terminal of the DUT. See
example.
4. Set the DUT on your test adapter.
5. Close the fixture cover and perform measurement.
N1259A Opt 013
Emi er / Source
Collector / Drain
A
Force
Sense
Force
Sense
2
1
5
3
4
Curve Tracer Test Adapter Socket
Internal connection
B
C
1: Collector/Drain Force
Base / Gate
2: Collector/Drain Sense
3: Emitter/Source Force
4: Emitter/Source Sense
Force
Sense
5: Base/Gate Force
6: Base/Gate Sense
Maximum voltage: 3000 V
6
Maximum current:
A-Force: 500 A pulse, 39 A dc
A-Sense: 40 A pulse, 2 A dc
B-Force: 500 A pulse, 39 A dc
B-Sense: 40 A pulse, 2 A dc
C-Force: 40 A pulse, 2 A dc
C-Sense: 40 A pulse, 2 A dc