Input Connection - Keysight Technologies B1505A Configuration Manual

Power device analyzer/curve tracer
Hide thumbs Also See for B1505A:
Table of Contents

Advertisement

Input Connection

Prepare the required cables listed in the following tables and connect them between the
B1505A terminals and the relative N1259A terminals.
without module selector, fast switch, and HVMCU.
Table 2-2
Table 2-3
(N1267A). For the additional information on the fast switch, see
HVSMU/HCSMU Fast Switch" on page
Table 2-4
information on the HVMCU, see
Figure 2-1
N1259A Rear View
Interlock
WARNING
The connector cap must be connected to the unused input connectors.
NOTE
For connecting HP/MPSMU, use either 16494A or 16493K.
For connecting MFCMU, use either AUX or MFCMU only for N1259A-020.
Keysight B1505A Configuration and Connection Guide
shows the additional connections for using module selector (N1259A-300).
shows the additional connections for using HVSMU/HCSMU fast switch
shows the additional connections for using HVMCU. For the additional
HPSMU1
HPSMU2
±200 V
Max
±200 V
Max
AUX
±200 V
Max
1
Force
Force
2
Sense
Sense
MF CMU Input
±25 V
Max
HPSMU3
±200 V
Max
Hcur
Hpot
Force
Lcur
Lpot
Sense
Table 2-1
5-32.
"To Connect HVSMU Current Expander" on page
Input
HVSMU1
GNDU1
HCSMU 1
±3 kV
Max
± 40 V
Max
Force
Sense
To avoid electrical shock and instrument damage,
do not remove the short cap and cables during operation.
Module Selector Input
HVSMU2
GNDU2
HCSMU 3
±3 kV
Max
± 40 V
Max
Force
Sense
To avoid electrical shock and instrument damage,
do not remove the short cap and cables during operation.
N1259A Connection Guide
Input Connection
shows the connections
"To Connect
5-34.
HCSMU2
±40 V
Max
Force
Sense
LINE
100– 240 V
Digital I/O
50/60 Hz
35 VA Max
N10149
2-3