Note For 4142B Users - Keysight Technologies B1505A Configuration Manual

Power device analyzer/curve tracer
Hide thumbs Also See for B1505A:
Table of Contents

Advertisement

Configuration Guide

Note for 4142B Users

Note for 4142B Users
The B1505A expands the maximum measurement range up to 40 A/3000 V from the range
10 A/1000 V covered by Keysight 4142B Modular DC Source/Monitor.
Because of the following reasons, the measurement setup and environments for the 4142B
cannot be used with the B1505A. The test fixture and the connection path from connector
plate to wafer prober must be designed, changed, or remade for the B1505A.
Table 1-32
Considering Connector Types
High current connector
High voltage connector
Interlock connector
1-74
Test fixture 16088A/B designed for the 4142B, cannot be used for the measurements
over 10 A/1000 V.
The connection path, connectors, cabling, and wiring will not support up to
40 A/3000 V.
Connector types are different from the 4142B.
BNC and screwed-triaxial
BNC female
bulkhead-type
Triax. female
bulkhead-type
HV Triax. female bulkhead-
type (N1254A-500)
Interlock receptacle: Agilent P/N:1252-1419
or Hirose, HR10A-7R-6S
4142B
Dual triaxial
BNC
Keysight B1505A Configuration and Connection Guide
Keysight
B1505A
BNC and triaxial
HV triaxial
6-pin

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents