Table 5-12
To Connect 16493S-020 Input
B1505A terminals
HCSMU
Force
used for
Sense
master
HCSMU
Force
used for
Sense
slave
GNDU
Table 5-13
To Connect 16493S-020 Output
1
Output terminals
Low
Force
Coaxial cable with BNC(m) connector. N1254A-503 can be used.
Sense
Coaxial cable with BNC(m) connector. N1254A-503 can be used.
High
Force
Coaxial cable with HV(plug) connector. N1254A-506 can be used.
Use coaxial cable to minimize affect of inductance.
Sense
Triaxial cable with HV(plug) connector. N1254A-505 can be used.
1. This is the DHCSMU channel. The connectors are compatible with the N1258A module selector output
connectors.
2. Non-Kelvin connection is not allowed. GNDU Force and Sense are internally connected to Low Sense.
Keysight B1505A Configuration and Connection Guide
Required cables
16493S HCSMU cable, 3 m or 1.5 m
16493S HCSMU cable, 3 m or 1.5 m
16493L GNDU cable, 3 m or 1.5 m
Required cables/Description
Connection Guide for Wafer Prober and Your Own Test Fixture
To Connect Dual HCSMU Adapter
Input terminals
HCSMU
Force
Master
Sense
HCSMU
Force
Slave
Sense
GNDU
Connect to
DUT low terminal
2
(ex: source)
DUT high terminal
2
(ex: drain)