Table 5-9
To Connect N1262A Output
Options
N1262A-001
Coaxial cable with SHV(plug) connector.
N1262A-002
Coaxial cable with SHV(plug) connector.
N1262A-010
Triaxial cable with Triaxial(m) connector.
N1262A-011
Coaxial cable with SHV(plug) connector.
N1262A-020
Triaxial cable with Triaxial(m) connector.
N1262A-021
Coaxial cable with SHV(plug) connector.
N1262A-023
Coaxial cable with UHV(plug) connector.
Voltage drop caused by the R-box must be subtracted from the measurement result as
following.
Vdut = Vout - Rs  Imeas
where, Vdut: Voltage after correction, Vout: Source output voltage, Rs: Resistance of the
series resistor, and Imeas: Measurement current

To Use Universal R-Box

To install your desired resistor, see the following simple instruction.
•tRequired parts:
Keysight B1505A Configuration and Connection Guide
Required cables/Description
wire, adequate rating, length, and quantity
your desired resistor, adequate rating
hexlobe (torx type) screwdriver T-10
label, for noting resistance value
Instruction:
1. Remove the cover of the universal R-box.
2. Solder the resistor and wire as shown below.
To prevent discharge and any accident, make enough space between the resistor
leads and the shield or chassis. For example, make space of about 1 mm from the
leads for maximum 200 V output, 6 mm for 3000 V, and 32 mm for 10 kV.
Case 1: N1262A-020/021, see
Connect a wire between a terminal and the center conductor of its near
connector.
Connect a wire between the other terminal and the center conductor of the other
connector.
Connect the resistor between the terminals.
Connection Guide for Wafer Prober and Your Own Test Fixture
Figure
5-1.
To Connect High Voltage R-Box