Table 2-4
To Connect between B1505A and N1259A for using HVMCU (N1266A)
B1505A terminals
16493L-001 GNDU cable, 1.5 m, 1 ea.
GNDU
Force
16493T-001 HVSMU cable, 1.5 m, 1 ea.
HVSMU
2
Force
16494A-001 Triaxial cable, 1.5 m, 1 ea.
MCSMU
Sense
16494A-001 Triaxial cable, 1.5 m, 1 ea.
2
Force
16494A-001 Triaxial cable, 1.5 m, 1 ea.
MCSMU
Sense
16494A-001 Triaxial cable, 1.5 m, 1 ea.
Digital I/O
16493G-001 Digital I/O connection cable, 1.5 m, 1 ea.
1. To configure HVMCU, the N1266A current expander is required.
2. HCSMU is substitutable. Then use 16493S HCSMU cable, N1254A-103 adapter for Force, and
N1254A-517 adapter for Sense.
N1266A terminals
Output
High
16493T-001 HVSMU cable, 1.5 m, 1 ea.
Low
16493L-001 GNDU cable, 1.5 m, 1 ea.
Digital I/O Output
16493G-001 Digital I/O connection cable, 1.5 m, 1 ea.
LINE
Power cable, 1 ea., for connecting to power line
1. This connection is required if the module selector (N1259A-300) is used.
Keysight B1505A Configuration and Connection Guide
Required cables
Required cables
N1259A Connection Guide
Input Connection
1
N1266A terminals
GNDU
HVSMU
Force
I Control
Sense
Force
V Control
Sense
Digital I/O Input
N1259A terminals
HVSMU1 or 2
GNDU1 or 2
1
Digital I/O
2-7