Controlling N1258A input-to-output path
The configuration and the default I/O path of the N1258A are set by using the Module
Selector tab screen of the Configuration window of the EasyEXPERT software. The
connection path from the input port to the output port is controlled by each test setup. And
the N1258A switching status is indicated by the Status indicator mounted on the front
panel. See
Figure 5-14
Module Selector Simplified Internal Connections
Table 5-22
Module Selector Status Indicator and Input-to-Output Connection Path
Measurement resource
used for the test
2
HPSMU or MPSMU
HCSMU or DHCSMU
HVSMU or HVMCU
HVSMU/HVMCU and
built-in series resistor
1. In the normal state, the N1258A makes the path to the output port from the input port specified by the
Default field. When a measurement is performed, the N1258A performs automatic switching in every
test. The N1258A makes the path to the output port from the measurement resource used for the test.
2. This condition is when all Input fields HVSMU, HCSMU, and HPSMU are blank.
Keysight B1505A Configuration and Connection Guide
Connection Guide for Wafer Prober and Your Own Test Fixture
Figure 5-14
and
Table 5-22
Low
S
F
P.A
F
S
S
GNDU
HCSMU
Input-to-output connection
Status
indicator
Low Sense
Open
HCSMU Low
Sense
HPSMU
+ GNDU Force
+ GNDU Sense
HCSMU
HVSMU
for the indication and the connection path.
High
S
F
100 k
P.A
F
F
S
F
HPSMU
HVSMU
1
of output terminals
Low Force
High Sense
HCSMU
Low Force
HPSMU
HCSMU
High Sense
HVSMU
HVSMU
+ Series
To Connect Module Selector
F: Force
S: Sense
P.A:
Protection adapter
GNDU force and sense
are connected to Low
sense line.
HVSMU force is connected
to High sense line.
High Force
Open
Open
HPSMU
Sense
Force
HCSMU
High Force
Force
Open
Force
resistor
5-31