Table 1-26
Accessories for Wafer Prober or Your Own Fixture
Model/Option
Module Selector
N1258A
Module Selector for B1505A
For High Voltage Capacitance Measurement
N1260A
High Voltage Bias-Tee
Keysight B1505A Configuration and Connection Guide
Accessories for Wafer Prober and Your Own Test Fixture
Description
Additional Information
The N1258A is furnished with 1.5 m
digital I/O control cable (16493G-001).
Maximum 65 VA
The N1258A has a built-in GNDU
protection adapter, a built-in HPSMU
protection adapter and a built-in
HCSMU Kelvin adapter.
Do not connect or put any conductor on
the HCSMU Low Force and Low Sense
terminals, outer conductor of the
coaxial connectors. Connecting or
putting conductor of circuit common,
chassis ground, or any potential on
causes the measurement error.
The N1260A-STD is a mandatory
option.
The N1260A High Voltage Bias-Tee is
not furnished with any cables.