Inline Package Socket module (N1265A-010)

This module provides a socket used for connecting three-terminal inline packaged device
and three couples of the Force and Sense terminals. Socket module internal connection is
shown in
Short bar is furnished with the module. It is used for performing the short correction before
the impedance measurement. Set the short bar before the short correction and remove it
after the correction.
CAUTION
Do not apply voltage/current over the maximum limit of the socket module.
Figure 3-7
Inline Package Socket Module
To use this module, see the following simple instruction.
1. Attach the socket module to the test fixture.
2. Connect wires between the socket module terminals and the fixture output terminals.
Keysight B1505A Configuration and Connection Guide
Figure
3-7.
165 mm
N1265A Opt 010
Emi er / Source
Collector / Drain
2
F
S
F
1
2
1
5
3
Inline Package Socket (3 Pin)
14 mm
3.48 mm
1.78 mm
Then use the following wire (furnished with the N1265A-010).
180 mm length wire (yellow) for Selector Output High/Low Force
180 mm length wire for Selector Output High/Low Sense
250 mm length wire for Gate, SMU, or chassis
For making the Kelvin connection, Force and Sense must be connected to Force and
Sense of the socket module respectively.
For the high voltage capacitance measurement, use SHV(plug)-SHV(plug) cable
(N1254A-512) and SHV(jack)-banana adapter (N1254A-513) for connection.
Selector Output and Gate should be connected as follows.
Selector Output High to DUT high (ex. Collector/Drain)
Selector Output Low to DUT low (ex. Emitter/Source)
Gate High to DUT drive (ex. Base/Gate Force)
Gate Low to DUT low (ex. Emitter/Source Force)
3
1
Base / Gate
S
F
S
2
3
165 mm
6
4
Height: 34 mm
(45 mm including terminal)
(58 mm including socket)
N1265A Connection Guide
Output Connection
Maximum voltage: 3000 V
Maximum current:
1-F: 500 A pulse, 39 A dc
1-S: 40 A pulse, 2 A dc
2-F: 500 A pulse, 39 A dc
2-S: 40 A pulse, 2 A dc
3-F: 500 A pulse, 39 A dc
3-S: 40 A pulse, 2 A dc
F: Force
S: Sense
3-17