To Connect Interlock Circuit - Keysight Technologies B1505A Configuration Manual

Power device analyzer/curve tracer
Hide thumbs Also See for B1505A:
Table of Contents

Advertisement

Figure 5-27
Dimensions of the Interlock Switch (Keysight N1254A-402)

To Connect Interlock Circuit

The B1505A provides the Interlock connector to prevent you from receiving an electrical
shock from high voltage (more than ± 42 V). If the interlock circuit is open, the B1505A
cannot apply high voltage more than ± 42 V.
Before performing measurement, connect the 16493J interlock cable between the B1505A
Interlock connector and the interlock connector which is a part of the interlock circuit
installed in your DUT interface as described in
Keysight B1505A Configuration and Connection Guide
Connection Guide for Wafer Prober and Your Own Test Fixture
2.8
10.3
4.3
2.8
35.6
Switch off
3.2
2.8
15.3
10.3
3.1
To Install an Interlock Circuit
6.4
2.8
8.1
NC
NO
C OM
22.2
14
27.8
"Procedure" on page
5-48.
4.75
3.1
6.3
15.9 18.8
Units: mm
5-49