Keysight Technologies B1505A Configuration Manual page 254

Power device analyzer/curve tracer
Hide thumbs Also See for B1505A:
Table of Contents

Advertisement

Connection and Ordering Examples
Configuration Examples for Lateral Device Measurement with Wafer Prober
Figure 7-19
Connection example for 3kV, 40A measurement without N1258A Module Selector
B1505A
For Drain
HVSMU
For Drain
F
HCSMU
S
F
For Drain
HCSMU
S
16494A
F
(Included in the MCSMU)
MCSMU
For Gate
S
F
For Chuck
MPSMU
S
GND
16494A
(Included in the MPSMU)
16493J Interlock cable (Included in B1505A)
NOTE
Module selector is available up to 30 A. If 40 A capability is required, please connect
16493S-020 Dual HCSMU Kelvin Connection Adapter without a N1258A Module
Selector (See
NOTE
Prober vendor is responsible for cabling inside the shielding box.
Table 7-17
Required equipment, accessories, and cables
Model/Option
Quantity
B1505A
B1511B-FG
B1512A-FG
B1513C-FG
B1514A-FG
16493S
16493S-020
16493S-021
16493S-011
16493T
16493T-001 or
002
16493U
7-36
16493T (Included in the HVSMU)
16493S (Included in the HCSMU)
16493S (Included in the HCSMU)
16493L
(Included in B1505A)
N1254A-104 TRX(J) to BNC(P) adapter
HCSMU Non-
Kelvin adapter
16493S-011
Protec on
Adapter
N1261A-003
Figure
7-19).
1
Power Device Analyzer/Curve Tracer mainframe
1
Medium Power Source/Monitor Unit module (100 V/ 100 mA)
2
High Current Source Monitor Unit, 20 A/20 V(Pulsed); 1 A/40 V(DC)
1
High Voltage Source Monitor Unit, 3000 V/4 mA (Pulsed & DC)
1
Medium Current Source Monitor Unit, 1 A/30 V(Pulsed), 100 mA/30 V(DC)
1
High Current Source Monitor Unit Cable
1
Dual HCSMU Combination Adapter (This adapter is used when a N1258A
Module Selector is not used)
1
Dual HCSMU Combination Adapter (This adapter is used with a N1258A
Module Selector)
1
High Current Source Monitor Unit non-Kelvin Adapter
1
High Voltage Source Monitor Unit Cable
2
High Voltage Source Monitor Unit Cable (1.5 m)
1
High Current BNC Coaxial Cable
R-Box
100kO
N 1262A-002
HF
Dual HCSMU
HS
Dual HCSMU
Combina on
adapter
Kelvin
LF
Combina on
16493S-021
adapter
16493S-020
LS
16493U-001
R-Box
1kO
N1262 A-010
16494A-003
16493U-001
1250-2405
16493T
Description
Keysight B1505A Configuration and Connection Guide
D
G
Chuck
S

Hide quick links:

Advertisement

Table of Contents
loading