Table 5-26
To Connect N1266A Input
B1505A terminals
Force
16493T-001 HVSMU cable, 1.5 m
HVSMU
16493L-001 GNDU cable, 1.5 m
GNDU
1
Force
16494A-001 Triaxial cable, 1.5 m
MCSMU
Sense
16494A-001 Triaxial cable, 1.5 m
1
Force
16494A-001 Triaxial cable, 1.5 m
MCSMU
Sense
16494A-001 Triaxial cable, 1.5 m
Digital I/O
16493G-001 Digital I/O connection cable, 1.5 m
Power cable, for connecting to power line
1. HCSMU is substitutable. Then use 16493S HCSMU cable, N1254A-103 adapter for Force, and
N1254A-517 adapter for Sense.
CAUTION
The GNDU can sink current of up to 4.2 A. Use 16493L GNDU cable to connect the
GNDU to the N1267A.
Do not use normal triaxial cable (16494A) because the maximum current rating of the
cable is 1 A.
Table 5-27
To Connect N1266A Output
N1266A
terminals
Low
16493L GNDU cable, N1261A-002 protection adapter and two
BNC(m) coaxial cables (N1254A-503 can be used), or N1261A-004
protection adapter and two SHV(plug) coaxial cables
Protection adapter is not required to connect to the N1258A,
N1259A, N1259A-300, or N1265A.
High
HV(plug) triaxial cable (N1254A-505 can be used).
CAUTION
Never connect the HVSMU Force and Guard terminals and the HVMCU High terminal to
any output, including circuit common, chassis ground, or any other measurement resource
such as SMU. Connecting other measurement resource may damage the connected one.
Keysight B1505A Configuration and Connection Guide
Connection Guide for Wafer Prober and Your Own Test Fixture
Required cables
Required cables/Description
To Connect HVSMU Current Expander
N1266A terminals
HVSMU
GNDU
I Control
V Control
Digital I/O Input
LINE
Connect to
DUT low terminal
(ex: source)
DUT high terminal
(ex: drain)
Force
Sense
Force
Sense
5-35