N1265A Connection Guide
Output Connection

To Connect DUT

WARNING
Set the instrument output off before connecting or disconnecting connection wire.
Press the B1505A front panel
the B1505A front panel High Voltage indicator is not lit.
If the N1268A is used, press the
output. And confirm that this red switch is not lit.
CAUTION
Do not connect or disconnect your DUT while the B1505A is applying voltage or current.
Otherwise, your DUT may be damaged.
CAUTION
Do not touch the terminals of the cable and wire. Oil, perspiration, and dirt prevent good
electrical contact, deteriorate insulation, and degrade measurement accuracy.
CAUTION
Never connect the Selector Output High Force, High Sense, and Guard terminals, Bias Tee
High connector, and UHV High connector to any output, including circuit common,
chassis ground, or any other measurement resource such as SMU. Connecting other
measurement resource may damage the connected one.
NOTE
SMU Guard terminals and Selector Output Guard terminal
Guard must be opened. You may extend it as close as possible to a DUT terminal for
reducing the leakage current of the extension cable.
1. Set the instrument output off.
2. Open the fixture cover.
3. Set a socket module on the test fixture.
4. Connect a wire to an output terminal. And repeat this for all terminals connected to the
5. Connect the wire to the socket module input terminal.
6. Set the DUT on the socket.
7. Close the fixture cover.
3-10
For using the blank silicon plate, see
3-22.
measurement resources to be used. See
Confirm the DUT pin assignment and the measurement resource to be used and
complete the connections of all measurement terminals without making misconnection.
For using the inline package socket module, also see
(N1265A-010)" on page
3-17.
For using the universal socket module, also see
(N1265A-011)" on page
3-18.
To apply more than ± 42 V, close the fixture cover. Otherwise, the interlock function
will stop the B1505A output over ± 42 V.
key to set the module output off. And confirm that
Stop
switch to disable the high voltage
High Voltage Enable
"Blank Silicon Plate (N1265A-002)" on page
Figure 3-2
for the output terminals.
"Inline Package Socket module
"Universal Socket Module