N1272A and N1273A Connection Guide
Input Connection
Figure 4-2
Selector Simplified Internal Connections
Figure 4-3
N1272A Rear View
1. Digital I/O connectors
2. Input
3. Output
4. LINE input receptacle and power switch
4-4
3
a
b
f
e
Output
Interlock
Collector/Drain
Base/Gate
Lcur
T2
T1
HVSMU
±3 k V
Max
±100 V
Max
Emitter/Source
AC/DC Guard
Lpot
T3
T4
±3 k V
±3 k
Max
Max
Force
Circuit
±3 k V
±100 V
Max
Max
Common
Guard
To avoid electrical shock and instrument damage,
do not connect/disconnect the cables during operation.
g
d
c
a. MFCMU input connectors Hcur, Hpot, Lpot, and Lcur
b. SMU input connectors
c. GNDU input connector
d. HVSMU input connector
e. Interlock input connector
a. Collector/Drain output connector
b. Base/Gate output connector
c. Emitter/Source output connector
d. AC/DC Guard output connector
e. HVSMU output connector
f. Interlock output connector
g. Frame ground terminal
2
a
b
c
e
d
Input
MFCMU
SMU
Force
Interlock
Hcur
GNDU
HVSMU
Sense
Hpot
±3 k V
Max
±25 V
Max
±100 V
Max
Keysight B1505A Configuration and Connection Guide
1
4
~ LINE
Digital I/O
100-240 V ~
Input
Output
50/60 Hz
70 VA Max