Reset Procedures; Warm Start - ABB NGC8200 User Manual

Gas chromatograph
Table of Contents

Advertisement

1.
In PCCU, go to the Operate drop-down menu.
 Select File Utilities then click the Save and Restore Utility.
 Or click the Save and Restore Utility on the toolbar.
2.
In the Save and Restore window, click Restore Station Files.
3.
When the Restore Station Files window appears, verify the default name and path for the files.
Click OK. This will restore the files to the tfCold drive.
IMPORTANT NOTE: It may not always be desirable to restore the station files to tfCold. Some
problems addressed in the
4.
Perform a cold start following the instructions in section 10.9.2, and verify that the device is
functioning properly.
10.9

Reset procedures

On occasion, it may be necessary to reset the device. There are two types of reset procedures: warm
or cold.
10.9.1

Warm start

A warm start occurs when the main power is removed and then re-applied while memory backup is
enabled. This does not clear the data stored in RAM. The warm start will only reset the analyzer's
microprocessor and not disturb any data that has been stored in RAM. A warm start can be used when
a power or communication interruption caused the microprocessor to lock up. There are two ways to
perform a warm start: using PCCU Entry view or pressing the reset button.
To perform a warm start with Entry view in PCCU:
1.
In PCCU Entry view, select the Station Setup tab.
2.
Locate the System shutdown / then Restart field.
3.
From the drop-down menu, select Yes.
Figure 10-9: Warm start using PCCU
4.
Click Send.
5.
Click Yes to confirm.
To perform a warm start with the reset button:
DANGER – Explosion risk. Serious damage to health / risk to life. Do not open or remove
covers unless the area is known to be non-hazardous.
Troubleshooting
section may require a selective restore.
21 01 51 0MN A H | NG C 82 0 0 A N D P GC 10 00 US ER MA NUA L | 1 0 5