Electrical Connections Analog Output Modules - ABB EasyLine Series Instructions For Installation Start-Up And Operation

Continuous gas analyzers
Table of Contents

Advertisement

84
EasyLine EL3000 Series Commissioning Instructions

Electrical Connections Analog Output Modules

Pin
Signal
1
AO1+
2
AO1–
3
AO2+
4
AO2–
5
AO3+
6
AO3–
7
AO4+
8
AO4–
Analog Outputs AO1 to AO4
0/4...20 mA (pre-set ex works to 4...20 mA), common negative pole,
electrically isolated to ground, can be connected to ground as required, in
this regard, max. gain compared to local protective ground potential 50 V,
working resistance max. 750 Ohm. Resolution 16 bit. The output signal
may not be less than 0 mA.
Design
4-pole or 8-pole plug-in terminal strip. Please refer to the information
about the requisite material (see page 89)!
Terminal Layout
An analog output is allocated in the sequence of the sample components
for each sample component. The sequence of the sample components is
documented in the analyzer data sheet and on the type plate.
NOTE
The assignment of the terminals can be changed in the configurator.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Easyline el3040Easyline el3020

Table of Contents