Avantra 36/44 Service Manual
Since the amount of chemicals pumped into the tanks is always 100 cc, the
system automatically calculates the time interval between each pulse as 100 cc x
24 hours divided by the preset value:
Interval Time =
All calculations base on a 50% density. Consequently, the basis of all values
displayed is 50% density.
2.4.2.3
Topping-up Replenishment
If a low level condition occurs after the system has been switched off on the
control panel for more than 30 minutes, the topping-up replenishment function
automatically comes on. This procedure pumps one litre of developer and one
litre of fixer into the baths. If this action offsets the low level condition, the
processor will be ready the next time it is switched on. If the action fails to
offset the low-level condition, the system goes to low level mode, jog mode
stops, and the low-level light flashes.
If the level is correct after topping up and a low level occurs again after 30 or
more minutes, the system again activated the topping-up replenishmen mode.
2.4.2.4
Manual Replenishment
Pressing the MAN REPL key releases replenisher manually. Each activation adds
100 cc of developer and fixer.
2-10 Functional Analysis
100 cc x 24 hours
Preset oxy stdb replenishment value
= hours