7-2-14. Running an MCS Transport Test

This procedure runs film through the MCS without exposing it. It can be used to check operation of the
film transport.
1. From the main menu of the SERVICE TOOL, select Diagnostics>Test Prints>MCS Transport.
2. Use the "Copies" dropdown window to select the number of sheets of film you wish to run.
3. Click [Run].
4. If you wish to stop the process before all sheets are run, click [Stop].

7-2-15. Obtaining a Diagnostics Summary

This procedure will provide status of any diagnostic currently running in the IMAGER. It will also show the
results of any diagnostic that has been run recently.
1. From the main menu of the SERVICE TOOL, select Diagnostics>Diag Summary to display "Current
Status" of the diagnostics.
Note
If a Diagnostic is currently running, the name of the test and the insurance number of the user who
requested it will display.