7-5-6-2.

Removing and Inspecting the FILM PICKUP MODULE

1. Remove the MODULE as instructed in procedure 4--6-1. (See Figure 7-15 below for identification of
components referenced in the following procedures.)
Film Surface Sensor S4
2. Check the CUP PLATE ASSEMBLY as follows (see Figure 7-16):
- a. Do the CUPS slide back and forth freely in their slots? If not, check alignment of the VACUUM
HOSES attached to the cups. If they are impeding movement of the CUPS, reposition the
hoses.
- b. Are the CUPS finger tight? It not, tighten them.
- c. Are the CUP SPRINGS overextended? If so, replace them.
- d. Does the CUP PLATE ASSEMBLY rotate freely on its pivots. If not, check and bend the pivot
flanges, as necessary, to align them.
Cup Springs
Cup Slider
Vacuum Hose
2005 July Rev. B
7F3318
Film Out Sensor S6
Cups Engaged Sensor S10
Figure 7-15. FILM PICKUP MODULE COMPONENTS
Figure 7-16. Components of the CUP PLATE ASSEMBLY
Cup Plate Assembly
Heel Assembly
Driver Assembly
Frame
Pickup Position Sensor S3
Cup Plate Pivot