8-5-81. 29-925: DENSITOMETER Offset Failure with Light Source On
LOCAL PANEL Message
Internal Diagnostic Failure:
Densitometer-- Alarm sounds
Summary
Before every print and during diagnostics, a reading of the Density Sensor is taken with the Light Source
turned on. If this reading is outside the specified range, this error is declared.
Result of Error
The IMAGER is unable to print film. It can accept and queue new jobs, but it will not initiate new films. All
films in the IMAGER will be reprinted. The error can be cleared only by restarting the IMAGER.
User Action
1. Restart the IMAGER.
2. If the error persists, call Service.
FE Action
Perform the DENSITOMETER reference level adjustment, paragraph 3-5.
2005 July Rev. B