DENSITOMETER LIGHT SOURCE ASSEMBLY

!
Caution
Any CIRCUIT BOARD that is replaced in the IMAGER must be returned to Kodak for rework or
disposed of properly. PRINTED CIRCUIT BOARDS contain lead.
1. De--energize the IMAGER and remove the DENSITOMETER MODULE (4-4-1).
!
Warning
When the POWER CORD is plugged in, hazardous voltages are present in some areas of the
IMAGER. These voltages can cause severe injury or death.
2. Disconnect the 2 ELECTRICAL PLUGS from the BOARD (Figure 4-13).
3. Remove 2 ATTACHING SCREWS and remove the LIGHT SOURCE BOARD.
4. After reassembly and reinstallation of the DENSITOMETER MODULE, adjust the light source
reference level as described in paragraph 3-5 in Adjustments.
4-16
Figure 4-13. Removing the LIGHT SOURCE BOARD
2
3
7F3318
8100- -72L
2005 July Rev. B