4-9-5.

Feed Roller Torsion Springs

1. Remove the SUPPLY CARTRIDGE.
!
Warning
When the POWER CORD is plugged in, hazardous voltages are present in some areas of the
IMAGER. These voltages can cause severe injury or death.
2. De-energize the IMAGER and disconnect the POWER CORD.
3. Remove the FEED ROLLER MODULE (4-9-1).
Note
Be aware that the 2 TORSION SPRINGS are not identical. For reference in reassembly, note the
configuration of the SPRINGS as you remove them.
4. For each TORSION SPRING: Remove the end of the SPRING from the CHASSIS RETAINER to
release tension (Figure 4-36).
5. For each TORSION SPRING:
a. Remove the SCREW securing the SPRING and pivot to the FEED ROLLER BRACKET.
b. Slide the SPRING, pivot free of the BRACKET.
c. Remove the SPRING.
Note
In reassembly, make sure the FLANGES of the PIVOTS (Figure 4-32) for the TORSION SPRINGS
are seated flat against the FEED ROLLER BRACKET.
Chassis
Retainer
5a
2005 July Rev. B 7F3318
5b
Figure 4-36. Removing the Feed Roller Torsion Springs
Section 4 - - Disassembly/Reassembly
4-41