4-6-5.

FILM PICKUP MOTOR (DCM2)

1. Remove the supply cartridge.
!
Warning
When the POWER CORD is plugged in, hazardous voltages are present in some areas of the
IMAGER. These voltages can cause severe injury or death.
2. De--energize the IMAGER and disconnect the POWER CORD.
3. Remove the FILM PICKUP MODULE (4-6-1).
4. Disconnect the ELECTRICAL PLUG (Figure 4-20).
5. Remove the SPRING from the MOTOR BRACKET.
6. Remove 2 SHOULDER SCREWS and WASHERS. (The WASHERS may fall off after the SHOULDER
SCREWS are removed.)
7. Remove the MOTOR.
2005 July Rev. B 7F3318
Figure 4-20. Replacing the PICKUP MOTOR