4-11-3. PLATEN FILM SENSOR (S6)
Disassembly
1. Remove the SUPPLY CARTRIDGE.
!
Warning
When the POWER CORD is plugged in, hazardous voltages are present in some areas of the
IMAGER. These voltages can cause severe injury or death.
2. De--energize the IMAGER and disconnect the POWER CORD.
3. Remove the IMAGING MODULE from the IMAGER (4-11-1).
4. Disconnect the ELECTRICAL PLUG from the SENSOR (Figure 4-42).
!
Caution
If the same SENSOR is to be replaced, be careful not to damage the MOUNTING CLIPS during
removal.
5. Gently rock the SENSOR back and forth to disengage the MOUNTING CLIPS.
Reassembly
6. When replacing a SENSOR, make sure that the MOUNTING CLIPS are securely seated in the sheet
metal.
7. Check that the SENSOR ACTUATOR ARM operates freely in its SLOT.
4-48
Figure 4-42. Replacing the PLATEN FILM SENSOR (S4)
7F3318
2005 July Rev. B