7. Check that "Status" is Pass and that the laser dynamic range is between 140 and 160.
Note
In some cases you will have to repeat the adjustment several times. As you do, note whether the
measured values are improving.
8. After the adjustment is completed, run a calibration print before running any other print.
!
Caution
If a print is initiated before calibration, 27-622 (Media LUT Non-Monotonic) will occur.
2005 July Rev. B
7F3318
Section 3 - - Adjustments
3-11