b. Set the "Copies" window at 1.
c. Click [Run].
6. As the leading edge of the film comes up into the PROCESSOR:
a. Insert a small strip of film under the COVER for the DENSITOMETER/EXIT SENSOR
(Figure 7-23), to lift the ACTUATOR of the SENSOR. (This will prevent a "Jam in Area 5" error.)
Leave the SENSOR actuated for 10 -15 seconds.
b. Grasp and remove the film before it contacts the DRUM (see Figure 7-24).
7. Examine the film for scratches.
If the film does not have scratches, the problem is on the exit side of the PROCESSOR (FILM
•
DIVERTER, FELT PAD, COOLING PLATE or DENSITOMETER turnaround). See paragraph 7-6-3.
If the film does have scratches, they are occurring in the FILM PLATEN or FILM TRANSPORT.
•
See paragraph 7-6-4.
Figure 7-23. Actuating the DENSITOMETER/EXIT SENSOR
2005 July Rev. B
7F3318
R
3
Remove 2 screws from each side
to pull ASSEMBLY out.
Insert film strip to lift ACTUA-
TOR of DENSITOMETER/ EXIT
SENSOR.
8100- -220L
Section 7 - - Troubleshooting
7-65