8-5-13. Error 10-003: Image Datapath Error in Delivery Server Subsystem
LOCAL PANEL Message
Delivery Server Image Datapath
Error -- Alarm sounds
Summary
The image datapath software of the Delivery Server Subsystem has failed. The error can be caused by a
number of problems, such as failure to create and initialize the MCSDIS software, failure of the PCIO
BOARD or driver to initialize, or a read/write failure on the PCIO BOARD.
Result of Error
The DRE is not functional.
User Action
1. Restart the IMAGER.
2. If the error persists, call Service.
FE Action
8-16
Log Message
Delivery Server Internal Error
DICOM Message
PRINTER DOWN
0--Fatal
7F3318
Severity
2005 July Rev. B