7-2-6.
Running the Mechanical Diagnostics
These diagnostics include tests for DOOR SOLENOID SOL1 and all MOTORS in the IMAGER except the
VACUUM PUMP and the following MOTORS that run continuously: the PROCESSOR MOTOR and the
OPTICS SPINNER MOTOR. The diagnostics simply turn on the device selected for a short period, then
turn it off and report "Pass" or "Fail".
Note
Descriptive data for all MOTORS is provided in Table 7-2, and locations of the MOTORS are shown
in Figure 7-2.
1. Set the SERVICE SWITCH in the Service position.
2. From the main menu of the SERVICE TOOL, select Diagnostics>MCS>Mechanical Diagnostics.
3. From the Select Device dropdown menu, select the device to be tested. These include:
Door Solenoid
•
Kicker Motor
•
Vertical Transport Motor
•
Optics Motors
•
Platen Motors
•
Tray Motors
•
4. From the Select Operation dropdown menu, select the type of test or a specific MOTOR if the device
selected includes more than one MOTOR. "Devices" which include more than one MOTOR are:
Optics Motors -- You can select the Optics Attenuator Motor or the Optics Translation Motor for
•
testing. Or you can choose to command the Translation Motor to move the OPTICS to the
shipping position.
Platen Motors -- You can select the Feed (Platen Roller) Motor or the Film Centering Motor.
•
2005 July Rev. B
7F3318
Section 7 - - Troubleshooting
7-11