7-2-6.

Running the Mechanical Diagnostics

These diagnostics include tests for DOOR SOLENOID SOL1 and all MOTORS in the IMAGER except the
VACUUM PUMP and the following MOTORS that run continuously: the PROCESSOR MOTOR and the
OPTICS SPINNER MOTOR. The diagnostics simply turn on the device selected for a short period, then
turn it off and report "Pass" or "Fail".
Note
Descriptive data for all MOTORS is provided in Table 7-2, and locations of the MOTORS are shown
in Figure 7-2.
1. Set the SERVICE SWITCH in the Service position.
2. From the main menu of the SERVICE TOOL, select Diagnostics>MCS>Mechanical Diagnostics.
3. From the Select Device dropdown menu, select the device to be tested. These include:
Door Solenoid
Kicker Motor
Vertical Transport Motor
Optics Motors
Platen Motors
Tray Motors
4. From the Select Operation dropdown menu, select the type of test or a specific MOTOR if the device
selected includes more than one MOTOR. "Devices" which include more than one MOTOR are:
Optics Motors -- You can select the Optics Attenuator Motor or the Optics Translation Motor for
testing. Or you can choose to command the Translation Motor to move the OPTICS to the
shipping position.
Platen Motors -- You can select the Feed (Platen Roller) Motor or the Film Centering Motor.