Kodak DryView 8150 Service Manual page 70

Laser imager
Hide thumbs Also See for DryView 8150:
Table of Contents

Advertisement

Service Manual
Type of MODALITY
Kodak and other CR
(Computed Radiogra-
phy) and DR (Digital
Radiography) Systems.
Also Workstations and
imaging devices that
output data to a linear--
in--density contrast
curve.
Fuji CR
CT (Computed
Tomography) or MRI
(Magnetic Resonance)
DSA (Digital Subtrac-
tion Angiography), in-
cluding
C--Arm and Digital
Fluoro
Note: * The number in bold should be the first choice when you are selecting parameter values.
2-50
Table 2-4. Recommended TFT Sets for MODALITIES
TFT Set
Table No.
(Con-
trast)*
WRKSTN2A 1
2
3
4
FCR302C4
10
11
12
13
14
15
VER693C30 5, 7, 8
VER713C0
6, 7, 9
Alternative 1
VER693C0
6, 7, 9
Alternative 2
Density
Smoothing
(Dmax)*
Type
3.1
5
3.0
2.9
2.8
2.6
5
2.7
2.8
2.9
3.0
3.1
3.0--3.1
15
2.9--3.1
15
2.9--3.1
15
Notes
Select Table number
(contrast) and Dmax in
the pairs shown. Do not
use with Fuji CR.(See
Fuji CR below.)
Table number (contrast)
and Dmax must be cho-
sen in the pairs shown.
The contrast curves are
approximately linear in
density for most of the
pixel value range, but
become relatively flat at
high (bright) pixel val-
ues. Designed for use
with Fuji CR only.
Table 7 provides good
grayscale range in all
CT and MRI studies.
Table 5 has less mid--
tone contrast. Table 8
has more mid--tone con-
trast.
VER713C0 provides
brighter image back-
ground than
VER693C0. Lower table
numbers have less
mid--tone contrast.
Higher have more.
VER693C0 provides
darker image back-
ground than
VER713C0. Lower
Table numbers have
less mid--tone contrast.
Higher have more.
7F3318
2005 July Rev. B