Running The Fru Diagnostics - Kodak DryView 8150 Service Manual

Laser imager
Hide thumbs Also See for DryView 8150:
Table of Contents

Advertisement

Service Manual
7-2-5.

Running the FRU Diagnostics

These diagnostics test the 4 peripheral Micro Controllers on the MCS MicroComm Bus: DENSITOMETER,
OPTICS, RF TAG, and TRAY (Feeder). Three operations can be run on each Micro Controller:
Micro Controller Diagnostics -- This is the same test as the diagnostic run at power on.
Micro Controller Reset -- This operation resets the selected Micro Controller.
Micro Communications -- The operation tests whether the MCS CPU can communicate with the
selected Micro Controller.
1. From the main menu of the SERVICE TOOL, select Diagnostics>MCS>FRU Diagnostics.
2. To run a test operation:
a. Select the type of Diagnostic operation: Micro Controller Diagnostics, Micro Controller Reset, or
Micro Communications.
b. Select the desired Micro Controller: Densitometer, Optics, RF Tag, or Tray.
Note
For the Micro Communications test or the Reset operation, you can also select the MCS Micro
Controller.
c. Click [Run] to start the test operation.
Micro Controller Diagnostics will run once and display "Pass" or "Fail".
Micro Communications tests will continue to cycle until you click [Stop]. A "Pass" or "Fail" will
be displayed for each test.
If you select Micro Controller Reset, you will lose your connection to the LOCAL PANEL and a
10-910 error will display. You will have to shut down and restart the IMAGER.
Note
After running a test, select "Click to show Result Data" to display the test data.