Densitometer/Exit Sensor (S9) - Kodak DryView 8150 Service Manual

Laser imager
Hide thumbs Also See for DryView 8150:
Table of Contents

Advertisement

4-4-2.

DENSITOMETER/EXIT SENSOR (S9)

1. Power down and remove the DENSITOMETER MODULE (4-4-1).
!
Warning
When the POWER CORD is plugged in, hazardous voltages are present in some areas of the
IMAGER. These voltages can cause severe injury or death.
2. Remove the SCREW on the right side of the SENSOR COVER BOX (Figure 4-12).
3. Loosen the SCREW on the left side of the SENSOR COVER BOX and pivot the COVER BOX over
and to the left to expose the SENSOR.
4. Disconnect the PLUG from the SENSOR CONNECTOR.
!
Caution
If the SENSOR is to be replaced, be very careful not to damage the clips during removal. Also be
careful not to damage the fragile ACTUATOR of the SENSOR during disassembly and reassembly.
5. Gently rock the SENSOR back and forth to disengage its CLIPS from the CHASSIS.
Figure 4-12. Removing the DENSITOMETER/EXIT SENSOR
2005 July Rev. B 7F3318
3
2
Section 4 - - Disassembly/Reassembly
4
8100- -71L
5
4-15