Power-On Settings and Service Functions

Param Softkey Menu

PARAM
Reset Softkey
RESET
Save Softkey
SAVE
Sensor Softkey
SENSOR
40
When the PARAM softkey is pressed, the function of the softkeys changes as
shown below. These options can be accessed without disconnecting the pulse
oximetry cable from the instrument.
.
The RESET softkey can be used if any settings stored in memory have been
changed from factory default values. If YES is pressed, the instrument sounds
three tones and the settings return to factory default values. When NO is pressed,
no changes are made to the settings stored in memory.
When adjustable values are changed from factory default, the SAVE softkey can
be used to preserve the settings as institutional power-on default values. Pressing
YES stores the current settings in memory. The instrument sounds three tones
indicating that the changes have been saved as power-on default values. The new
saved values will continue to be used through power-on and off cycles until they
are changed and saved again, or until they are reset. If NO is pressed, the changed
values will not be saved.
Note:
An invalid tone indicates that a parameter value cannot be saved as a
power-on default. See Power-On Settings on page 35. Along with the
invalid tone, a message will be displayed indicating which parameter
could not be saved as a power-on default.
The SENSOR softkey enables/disables the Sensor Event Record function.