Specifications
94
Item
N-595 power entry module fuse holder
N-595 exterior markings
Front panel and case labeling
N-595 button spacing
Year of manufacture symbol
Conductive coating and polymeric
materials
Operation during physical shock
Operation during vibration
Electromagnetic Compatibility
Radiated and conducted emissions
Harmonic emissions
Voltage fluctuations/flicker emissions
Electrostatic discharge immunity
Radiated radio-frequency
electromagnetic field immunity
Electrical fast transient/burst
immunity
Surge immunity
Conducted EMI susceptibility
Power frequency magnetic fields
Operation with line voltage variations
Operation with electrical line voltage
variations
Radiated magnetic field emissions
Compliant With
IEC 60601-1, sub-clause 59.3
IEC 60601-1, sub-clause 6.1, 6.3, and 6.4;
EN 865, clause 6
IEC 60878, EN 980, ISO 7000, EN 60417-1,
EN 60417-2
ISO 7250
EN 980
UL 2601-1, clause 55
IEC 60068-2-27 at 100 g
IEC 60068-2-6 and IEC 60068-2-34
IEC 60601-1, sub clause 36, IEC/EN
60601-1-2 (second edition)
EN 55011, Group 1, Class B
IEC 61000-3-2
IEC 61000-3-3
EN 61000-4-2, level 3 table top equipment
IEC 61000-4-3 at 3V/m
IEC 61000-4-4, level 3
IEC 61000-4-5, level 3; FDA Reviewer's
Guide
IEC 61000-4-6 at 3 V/m
IEC 61000-4-8 at 3 V/m
IEC 61000-4-11 for Table 7
FDA Reviewer's Guide
RE 101/Army/7cm of MIL-STD-461E