Trend Data Printout (Ascii Mode); Trend Printout (Graph Mode); Figure 18: Trend Data Printout (Ascii Mode) - Nellcor OxiMax N-595 Service Manual

Hide thumbs Also See for OxiMax N-595:
Table of Contents

Advertisement

Trend Data Printout (ASCII Mode)

Trend Printout (Graph Mode)

N-595
Table 19: Operating Status Codes
Code
PS
SD
SH
SL
- - -
*
Note:
A Sensor Disconnect will also cause three dashes (- - -) to be displayed in
the patient data section of the printout.
The format of data displayed when a trend printout is requested is similar to that
of the real-time data. The only differences are that "TREND" is displayed in the
top row instead of the "CRC:XXXX" software verification number, and there is
no "Status" column (Figure 18).
Readings are displayed in 2-second intervals. The values on each row are an
average for the 2-second period.
At the end of the printout, an "Output Complete" line indicates that the
transmission was successful. If the "Output Complete" line is not present, the data
should be considered invalid.

Figure 18: Trend Data Printout (ASCII Mode)

The graph mode (Figure 19 and Figure 20) disables all printout functions except
trend data. Trend printouts will be graphical if connected to a serial printer that
Meaning
Pulse Search
Sensor Disconnect
Saturation Upper Limit Alarm
Saturation Lower Limit Alarm
No Data Available
Alarm Parameter Being Violated
Data Port Interface Protocol