Nellcor OxiMax N-595 Service Manual page 67

Hide thumbs Also See for OxiMax N-595:
Table of Contents

Advertisement

N-595
Error
Meaning
Code
264
SpO
back end reports corrupted variable
2
265
SpO
back end reports memory overflow
2
266
SpO
back end reports bad pointer
2
267
SpO
back end reports parameter value out-of-range
2
268
SpO
back end reports reset detected
2
269
SpO
back end reports unexpected value
2
270
SpO
back end reports time-out
2
271
SpO
back end reports not ready/not initialized
2
272
SpO
back end reports double fault
2
273
SpO
back end reports date out-of-range error
2
274
SpO
back end reports incompatible software version
2
275
SpO
back end reports incorrect registration number
2
276
SpO
back end reports sensor read failure
2
277
SpO
back end reports sensor signature verification fails
2
278
SpO
back end reports warmed sensor temperature set point failure
2
279
SpO
back end reports warmed sensor/SpO
2
280
SpO
back end reports does not support feature required by sensor
2
281
SpO
back end reports overflow/underflow
2
282
SpO
back end reports sensor activation failure
2
512
General failure of UIF Module generic post
512
Dead battery/Missing battery
514
Real time clock is non-operational
515
Application code is not present in the flash
516
Invalid flash type
517
Serial clock line is not toggling or is toggling at an incorrect rate
518
Application program is corrupt
519
Invalid FE102 version
520
Error in the start up sequence
521
OS multitasking service failure
522
A state machine has received an unknown state transition
Table 10: Error Codes
2
Troubleshooting
front end incompatible
59