Operation with a Live Subject
N-595
6. After the monitor completes its normal power-up sequence, verify that the
O
-M
sensor LED is brightly lit.
XI
AX
7. Slowly move the O
XI
element of the O
-M
XI
the LED approaches the optical O
decreases.
8. Open the O
M
sensor and notice that the LED intensity increases.
XI
AX
9. Repeat step 7 and the intensity will again decrease. This variation is an
indication that the microprocessor is in proper control of LED intensity.
10. Press the ON/STANDBY button to turn the monitor off.
Patient monitoring involves connecting the O
qualitative test.
.
1. Ensure that the monitor is connected to an AC power source.
2. Connect a DOC-10 pulse oximetry cable to the monitor sensor port.
3. Connect a Nellcor O
oximetry cable.
4. Clip the MAX-A to the subject as recommended in the O
directions for use.
-M
sensor LED in proximity to the photo detector
AX
sensor (close the O
AX
XI
M
sensor, that the LED intensity
XI
AX
M
XI
1 Sensor Port
M
MAX-A oxygen O
XI
AX
Performance Verification
-M
sensor slowly). Verify, as
AX
sensor to a live subject for a
AX
M
sensor to the pulse
XI
AX
M
sensor's
XI
AX
25