Power-On Settings and Service Functions

Accessing the Service Functions

LIGHT
38
All service functions are accessible when the DOC-10 pulse oximetry cable is
disconnected from the monitor. Disconnect the O
extension cable; or, disconnect the MC-10 extension cable from the instrument.
1. Turn on the N-595 by pressing the ON/STANDBY button.
2. Wait for monitor power-on self-test to complete.
3. Simultaneously press and hold the LIGHT softkey and the CONTRAST
button until the service softkeys are displayed.
Note:
The service function is only accessible from the main menu display. The
menu bar will change to the headings listed below.
Note:
If the above step is performed with a pulse oximetry cable connected,
only the PARAM and EXIT softkeys appear on the screen.
The following list can be used as a quick reference showing how to reach different
softkey functions. Items reached through the PARAM softkey can be accessed
during normal operation. Functions provided by the PRINT and NEXT softkeys
cannot be accessed when a pulse oximetry cable is connected to the instrument.
Each of the various functions is described in the text that follows.
PARAM (Service Menu)
-
RESET
-
-
RESET DEFAULTS?
-
-
-
YES (resets parameters to factory defaults, sounds three tones to
indicate that defaults have reset)
-
-
-
NO (back to Service menu)
-
SAVE
-
-
SAVE DEFAULTS?
-
-
-
YES (saves parameters as default settings, sounds three tones to
indicate that defaults have been saved)
-
-
-
NO (back to Service menu)
-
SENSOR (enables/disables sensor trend writing on rewritable sensors)
-
-
BACK (back to Service menu)
M
sensor from the MC-10
XI
AX