Nellcor OxiMax N-595 Service Manual page 23

Hide thumbs Also See for OxiMax N-595:
Table of Contents

Advertisement

EXIT
LIMITS
SELECT
EXIT
LIMITS
N-595
6. Press and hold the ADJUST UP button and verify that the %SpO
limit display cannot be raised past the upper alarm limit setting of "85."
7. Press the EXIT softkey.
8. Press the LIMITS softkey
9. Press the SELECT softkey three times. Verify that the monitor emits a beep
after each keystroke. The Pulse upper alarm limit should be "170" and should
be boxed.
10. Press and hold the ADJUST DOWN button.
11. Verify that the minimum displayed value is "41" for the BPM upper alarm
limit.
12. Press the EXIT softkey.
13. Press the LIMITS softkey.
Performance Verification
lower alarm
2
15