SETUP
NEXT
LANG
BACK
Nurse Call Setup
SETUP
NEXT
NEXT
N-595
2. Press the SETUP softkey.
3. Press the NEXT softkey.
4. Press the LANG softkey.
5. Use the ADJUST UP and ADJUST DOWN buttons to select the desired
language.
6. Press the BACK softkey to save the language setting.
The voltage polarity for the Nurse Call, available at pins 11 and 5, can be selected
through the softkeys. NORM + sets the voltage to +5 VDC to +12 VDC and
NORM - sets the voltage to -5 VDC to -12 VDC when there is no audible alarm.
When an audible alarm occurs, these voltages switch polarity. This signal is
available only if the instrument is operating on AC power. For more information,
see Nurse Call on page 122.
1. Turn on the N-595 by pressing the ON/STANDBY button.
2. Press the SETUP softkey.
3. Press the NEXT softkey.
4. Press the NEXT softkey.
Data Port Interface Protocol
109