2. The monitor pulse amplitude waveform will initially increase in amplitude
and then stabilize.
3. The monitor will display:
75 %SpO
(test pass criteria is 73 to 77 %SpO
2
60 BPM (test pass criteria is 57 to 63 BPM inclusive)
alarm
pulse amplitude indicator - high level modulation
4. Perform Test #1: BPM on page 28. The pulse amplitude indicator should
indicate high level modulation.
5. Perform Test #2: SpO2 on page 29. The pulse amplitude indicator should
indicate high level modulation.
6. Press the SRC-MAX % Modulation selection button. The SRC-MAX %
Modulation
LED will light.
7. The monitor pulse amplitude waveform will decrease in amplitude.
Performance Verification
inclusive)
2
31