Table 10: Error Codes - Nellcor OxiMax N-595 Service Manual

Hide thumbs Also See for OxiMax N-595:
Table of Contents

Advertisement

Troubleshooting
58
Table 10 provides a list of error codes for the N-595.
Error
Meaning
Code
1
SpO
front end RAM error
2
2
SpO
front end ROM/code integrity error.
2
3
SpO
front end reported a bad CRC
2
4
SpO
front end reported FSP message not allowed
2
5
SpO
front end reported illegal value sent in FSP messageSpO
2
6
SpO
front end reports calibration (offset) failure
2
9
SpO
front end reported syntax error in FSP message
2
10
Over-current limit in SpO
11
SpO
front end reports incorrect system voltage
2
12
SpO
front end reports other hardware problem
2
14
SpO
front end reports communication channel overflow
2
16
SpO
front end reports watch dog time out
2
17
SpO
front end reports that sensor appears defective
2
18
SpO
front end reports internal register appears modified from expected
2
value
19
SpO
front end reports signal out-of-range
2
48
SpO
front end reports spurious interrupt
2
49
SpO
front end reports internal buffer overflow
2
50
SpO
front end reports intermittent error
2
51
SpO
front end reports digital communications error
2
52
SpO
front end reports warmer error
2
53
Front end data not received
256
SpO
back end reports beginning of packet missing
2
257
SpO
back end reports packet start ID (SID) missing
2
258
SpO
back end reports packet length error
2
259
SpO
back end reports message length error
2
260
SpO
back end reports packet contains unsupported Key
2
261
SpO
back end reports packet CRC error
2
262
SpO
back end reports end of packet missing
2
263
SpO
back end reports packet contains undefined key
2

Table 10: Error Codes

front end has tripped
2
front end
2

Advertisement

Table of Contents
loading