Performance Verification
Test #1: BPM
28
4. After the monitor completes POST, the monitor will:
be in SpO
alarm
2
display an %SpO
display a pulse rate of 60 (pass criteria is 57 to 63 BPM inclusive)
pulse amplitude indicator - display low level modulation (low amplitude
pulse amplitude indicator)
1. Press the SRC-MAX % Pulse Rate selection button. The SRC-MAX Pulse
Rate 200 LED will light.
2. The monitor BPM will increase to 200 and stabilize at 200 BPM. The test pass
criteria is 197 to 203 BPM inclusive.
3. The monitor will display:
75 %SpO
2
200 BPM (pass criteria is 197 to 203 BPM inclusive)
alarm
pulse amplitude indicator - low level modulation
of 75 (pass criteria is 73 to 77 %SpO
2
inclusive)