Removal

70
1. Follow the procedure in paragraphs Prior to Disassembly on page 64, and
Monitor Disassembly on page 65.
2. Push the top of the Power Entry Module (PEM) in from the outside of the
case, and lift up.
3. Use needle-nose pliers to disconnect the leads from the PEM (see Figure 8).
4. Disconnect the fan wire harness from J1 on the Power Supply PCB (see
Figure 9).
5. Use a 10-mm wrench to disconnect the Power Supply ground lead from the
equipotential terminal (see Figure 8).
6. Remove the seven screws shown in Figure 9.