PRINT
N-595
Accessing the PRINT softkey makes four printouts available. See Data Port
Interface Protocol on page 105, for information about how to make connections to
the data port and how data is presented in a printout. The appropriate printout can
be selected by pressing the corresponding softkey. The softkey configuration that
appears after the PRINT softkey has been pressed is shown below.
Up to 48 hours of trend data can be viewed on the printouts described below.
When the monitor is turned on, trend data is recorded every 4 seconds. As an
example, an instrument that is used 6 hours a week would take approximately 8
weeks to fill its memory.
Note:
The two-letter codes and the symbols that occur in the printout are
described in Table 19 on page 120.