Power-On Settings and Service Functions
Trend Softkey
TREND
ERRLOG Softkey
ERRLOG
42
A Trend printout will include all data recorded for up to 48 hours of monitoring
since the last Delete Trends was performed. A new trend point is recorded every 4
seconds. The figure below is an example of a Trend printout.
The first row of the printout includes information about the type of instrument
delivering the information, the software level, type of printout, and alarm
parameters. The second line lists the headings for the columns. These lines are
printed out every 25 lines, or when a change to an alarm limit is made.
Patient data is represented with a date and time stamp for the data. In the example
above, the "- - -" means that an O
quality of the data being received was too low for the monitor to interpret the data.
Patient data that is outside of an alarm limit is marked with an asterisk (*).
At the end of the printout "Output Complete" will be printed. This indicates that
there was no corruption of data. If the Output Complete statement is not printed at
the end of the printout, the data must be considered invalid.
This softkey is for Nellcor's Customer Service Engineering Only.
A list of all the errors recorded in memory can be obtained by pressing the
ERRLOG softkey. The first line lists the type of instrument producing the
printout, software level, type of printout, and the time of the printout. The second
M
sensor was connected but the signal
XI
AX