Disassembly Guide

Prior to Disassembly

Fuse Replacement

64
that was installed when you started troubleshooting) and continue troubleshooting
as directed in this section.
1. Turn the N-595 off by pressing the ON/STANDBY button.
2. Disconnect the monitor from the AC power source.
1. Complete the procedure in paragraph Prior to Disassembly on page 64.
2. Disconnect the power cord from the back of the monitor.
3. Remove the fuse drawer from the power module by pressing down on the tab
in the center and pulling out as shown in Figure 4.
4. Put two new, 5 x 20-mm, slow blow, 0.5-amp, 250-volt fuses in the drawer
and reinsert the drawer in the power entry module.

Figure 4: Fuse Removal