Performance Verification
LIMITS
SELECT
14
The steps for changing an alarm limit are the same if the view being used is Blip
(Magnified) view.
Note:
Power-on defaults will be the factory-set defaults or the defaults set by
your institution.
1. Turn on the N-595 by pressing the ON/STANDBY button.
2. Press the LIMITS softkey. Verify that the monitor emits a single beep and the
pleth view is replaced with a display of the alarm limits. The upper alarm limit
for %SpO
will indicate an alarm limit of "100" (or institutional default
2
setting) inside a box.
3. Press and hold the ADJUST DOWN button. Verify that the boxed number for
%SpO
upper alarm limit reduces to a minimum of "86."
2
Note:
A decimal point in the display indicates that the alarm limits have been
changed from factory default values.
4. Press the SELECT softkey. Verify that the monitor emits a single beep and the
box moves to the %SpO
default setting).
5. Press and hold the ADJUST DOWN button and verify that the %SpO
alarm limit display reduces to a minimum of "20."
lower alarm limit of "85" (or your institutional
2
lower
2