Performance Verification

Battery Charge

Power-Up Performance

Power-On Self-Test
12
Perform the following procedure to fully charge the battery.
1. Connect the monitor to an AC power source.
2. Verify that the monitor is off and that the AC Power/Battery Charging
indicator is lit.
3. Charge the battery for at least 14 hours with the monitor turned off or 18 hours
with the monitor turned on.
The power-up performance tests verify the following monitor functions:
Power-On Self-Test on page 12
Power-On Defaults and Alarm Range Limits on page 13
1. Connect the monitor to an AC power source.
2. Verify that the monitor is off and that the AC Power/Battery Charging
indicator is lit.
3. Do not connect any cables to the monitor.
4. Turn on the N-595 by pressing the ON/STANDBY button. Observe the
monitor front panel. The monitor must perform the following:
Within 2 seconds all LEDs are illuminated, all pixels on the LCD display
are illuminated, and the backlight comes on.
The indicators remain lighted.
The LCD display shows NELLCOR and the software version of the
N-595.